Method 6010D Water

Metals by Inductively Coupled Plasma-Atomic Emission Spectrometry


This method is for the determination of trace metals from aqueous matrices by inductively coupled plasma atomic emission spectrometry (ICP).   An aliquot of the sample is solubilized or digested using with acid following the appropriate SW-846 preparation methods (3000 series methods).  See Chapter 3 of SW-846.  

If the analysis is for dissolved metals then acid digestion is not necessary if the sample ha been filtered and acid preserved prior to analysis.  The filtering for dissolved metals should be done immediately after sample collection.

NOTE:  This method updates and replaces Methods 6010, 6010A, and 6010B.


(EPA SW-846)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   6 months from sample collection to analysis for all elements except Hg which is 28 days from sample collection to analysis
Preservatives:   Add HNO3 to pH < 2.
Required Preps:   500mL Glass, HDPE or PTFE bottle
Collection Method:   Grab sampling
Analytical Methodology:   ICP/AES
Documentation:   6010D

Analyte List*

Analyte Formula CAS Number Detection Limit
Aluminum
Al
7429-90-5
30
 µg/L
Iron
Fe
7439-89-6
4.1
 µg/L
Lead
Pb
7439-92-1
28
 µg/L
Lithium
Li
7439-93-2
2.8
 µg/L
Magnesium
Mg
7439-95-4
20
 µg/L
Manganese
Mn
7439-96-5
0.93
 µg/L
Mercury
Hg
7439-97-6
0.007
 mg/L
Molybdenum
Mo
7439-98-7
5.3
 µg/L
Nickel
Ni
7440-02-0
10
 µg/L
Potassium
K
7440-09-7
15
 µg/L
Silver
Ag
7440-22-4
4.7
 µg/L
Sodium
Na
7440-23-5
19
 µg/L
Strontium
Sr
7440-24-6
0.28
 µg/L
Thallium
Tl
7440-28-0
27
 µg/L
Tin
Sn
7440-31-5
17
 µg/L
Titanium
Ti
7440-32-6
5
 µg/L
Antimony
Sb
7440-36-0
21
 µg/L
Arsenic
As
7440-38-2
35
 µg/L
Barium
Ba
7440-39-3
0.87
 µg/L
Beryllium
Be
7440-41-7
0.18
 µg/L
Boron
B
7440-42-8
3.8
 µg/L
Cadmium
Cd
7440-43-9
2.3
 µg/L
Chromium
Cr
7440-47-3
4.7
 µg/L
Cobalt
Co
7440-48-4
4.7
 µg/L
Copper
Cu
7440-50-8
3.6
 µg/L
Vanadium
V
7440-62-2
5
 µg/L
Zinc
Zn
7440-66-6
1.2
 µg/L
Calcium
Ca
7440-70-2
6.7
 µg/L
Silica (SiO2)
SiO2
7631-86-9
17
 µg/L
Phosphorus
P
7723-14-0
51
 µg/L
Selenium
Se
7782-49-2
50
 µg/L

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.