Method 0031/5041A/8260B

Volatile Organic Compounds from Tenax®/Tenax®/Anasorb®-747 Tubes by the SMVOC Sampling Method


These methods describe the sampling (0031), desorption (5041A), and analysis (8260B) of volatile organic compounds collected from stack gas effluents using volatile organic sampling train (Super-VOST) methodology.  This method is also referred to as the SMVOC method (sampling method for volatile organic compounds).  Samples are analyzed by thermal desorption of the sorbent tubes followed by GC/MS analysis.  For this method volatile organic compounds are defined as those compounds with boiling points in the range of -15°C to 121°C.  Although this method is not applicable for polar, water soluble and reactive volatile organic compounds. 

Samples are collected using a SMVOC sampling train consisting of a heated probe, water cooled condensers, Tenax®/Tenax®/Anasorb®-747 sorbent tubes, an empty impinger for condensate collection, and a silica gel drying tube.  Isokinetic conditions are not required for sample collection since the compounds of interest are in the vapor phase at the point of sample collection.  The upper concentration limit for volatile organic compounds in the effluent stream is 1.5ppm per the method. 

20L is the recommended sampling volume for collection and there are two recommended sampling rates referred to as FAST-VOST and SLO-VOST.  The sampling rate is 1 L/min. by the FAST-VOST method and 0.25 L/min. for 20 min. or 0.5 L/min. for 40 min. for the SLO-VOST method.  The goal of either method is to collect a total number of sorbent tube sets to encompass a total sampling time of 2 hours (one ‘run’): i.e., if a sampling rate of 1L/min for 20 minutes is used, a total of six sorbent tube sets will be collected in 2 hours of sampling.

When sampling for a compliance project, 3 runs as described above are collected  Analysis of the condensate collected in the empty impinger of the sampling train is also typically required for compliance projects.   The two Tenax® tubes typically will be analyzed in tandem with the Anasorb®-747 tube analyzed separately, although all three tubes may be analyzed individually or in tandem per guidance from the overseeing regulatory body.  Lower detection limits are achieved by combining all the tubes for analysis.  The method also states that it's acceptable to collect the condensate at the end of each 2 hour run. 

The method also requires the collection and analysis of a field blank and trip blank per sampling event.  A trip blank consisting of a set of Tenax®, Tenax®, Anasorb®-747 tubes from the same lot number and kept under the same conditions from the time the tubes leave the laboratory until they return to check for contamination from shipping and handling.

High moisture sources may cause problems with the analyses as the moisture will dampen the Tenax® and  Anasorb®-747 sorbents.  It is difficult to purge the POHCs from wet sorbents resulting in poor desorption efficiencies.

NOTE: SW-846 Method 5040 has been replaced by SW-846 Method 5041A, and SW-846 Method 8240 has been replaced by SW-846 Method 8260B.


(EPA SW-846)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   14 days from sample preparation to collection and 14 days from sampling to analysis
Preservatives:   Samples kept at less than 10°C for storage and shipment to the laboratory
Required Preps:   Tenax®/Tenax®/Anasorb®-747 tubes
Collection Method:   SMVOC train following SW-846 Method 0031 procedures.
Analytical Methodology:   Thermal desorption GC/MS (low resolution) by SW-846 Method 8260B
Documentation:   0031
5041A
8260B

Analyte List*

Analyte Formula CAS Number Detection Limit
1,2-Dichloroethane
C2H4Cl2
107-06-2
0.1
 µg total catch weight
Acrylonitrile
C3H3N
107-13-1
0.1
 µg total catch weight
Toluene
C7H8
108-88-3
0.1
 µg total catch weight
Chlorodibromomethane
CHBr2Cl
124-48-1
0.1
 µg total catch weight
Chloroprene
C4H5Cl
126-99-8
0.1
 µg total catch weight
Tetrachloroethene
C2Cl4
127-18-4
0.1
 µg total catch weight
trans-1,2-Dichloroethene
C2H2Cl2
156-60-5
0.1
 µg total catch weight
1,3-Dichloropropene
C3H4Cl2
542-75-6
0.1
 µg total catch weight
Carbon tetrachloride
CCl4
56-23-5
0.1
 µg total catch weight
Chloroform
CHCl3
67-66-3
0.1
 µg total catch weight
Benzene
C6H6
71-43-2
0.1
 µg total catch weight
1,1,1-Trichloroethane
C2H3Cl3
71-55-6
0.1
 µg total catch weight
Dibromomethane
CH2Br2
74-95-3
0.1
 µg total catch weight
Methylene chloride
CH2Cl2
75-09-2
0.1
 µg total catch weight
Carbon disulfide
CS2
75-15-0
0.1
 µg total catch weight
Bromodichloromethane
CHBrCl2
75-27-4
0.1
 µg total catch weight
1,1-Dichloroethane
C2H4Cl2
75-34-3
0.1
 µg total catch weight
1,1-Dichloroethene
C2H2Cl2
75-35-4
0.1
 µg total catch weight
Trichlorofluoromethane
CCl3F
75-69-4
0.1
 µg total catch weight
1,2-Dichloropropane
C3H6Cl2
78-87-5
0.1
 µg total catch weight
1,1,2-Trichloroethane
C2H3Cl3
79-00-5
0.1
 µg total catch weight
Trichloroethene
C2HCl3
79-01-6
0.1
 µg total catch weight

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.