Method 8

Vinyl bromide


This method is for the determination of vinyl bromide. The sample is collected charcoal sorbent tubes, desorbed with carbon disulfide, and then analyzed by gas chromatography with a flame ionization detector. Samples should be stored in a refrigerator and analyzed as soon as possible.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   Analyze as soon as possible.
Preservatives:   Samples must be refrigerated.
Required Preps:   Solid sorbent tubes: coconut shell charcoal, 100mg/50mg, 7cm x 6mm OD x 4mm ID.
Collection Method:   Solid sorbent charcoal tubes: 5 L @ 0.2 L/min.
Analytical Methodology:   GC/FID
Documentation:   008

Analyte List*

Analyte Formula CAS Number Detection Limit
Vinyl Bromide
C2H3Br
593-60-2
4
 ng

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.