Method 1

Cyclohexanone by GC/FID


This method is for the determination of cyclohexanone. Sample vapors are collected on chromosorb 106 tubes, desorbed with carbon disulfide, and analyzed by gas chromatography with flame ionization detection. The cyclohexanone collected on Chromosorb 106 remains stable for at least 16 days, even at room temperature.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   The recovery of the analyte from the collection medium after storage must be 75% or greater. Recoveries of greater than 89% were obtained from samples stored for up to 16 days at both ambient (23°C) and refrigerated (-5°C) temperatures.
Preservatives:   Store refrigerated or at room temperature.
Required Preps:   Chromosorb 106 tube: 100 mg/50 mg, 8.5cm x 6mm OD x 4mm ID
Collection Method:   Chromosorb 106 tube: 10 L @ 0.05 - 0.2 L/min.
Analytical Methodology:   GC/FID
Documentation:   001

Analyte List*

Analyte Formula CAS Number Detection Limit
Cyclohexanone
C6H10O
108-94-1
0.8
 ng

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.