Method ID-111

Phosphoric Acid in Workplace Atmospheres


This method is for the collection and analysis of phosphoric acid in workplace air.  The sample is collected using a mixed cellulose ester membrane (MCE) filter.  After collection, the filter is desorbed with a sodium carbonate/sodium bicarbonate solution and analyzed by ion chromatography (IC).


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   None specified in the method.
Preservatives:   None specified in the method.
Required Preps:   37mm MCE filter with a 0.8 micron pore size (SKC 225-3-01)
Collection Method:   The method recommends a 2 L/min. flow rate with a 960L sample collection volume.
Analytical Methodology:   IC
Documentation:   ID-111

Analyte List*

Analyte Formula CAS Number Detection Limit
Phosphoric acid
H3PO4
7664-38-2
5
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.