Method 7602 is used for the determination of quartz. Samples are collected using a cyclone and filter and analyzed using infrared absorption spectrophotometry. NOTE: This is P&CAM 110. It is similar to Method 7603, except for sample preparation (KBr pellet vs. redeposition).
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | N/A |
Preservatives: | N/A |
Required Preps: | 37 mm, 0.8 µm MCE, 5.0 µm PVC filter |
Collection Method: | Cyclone and filter, 400-800 L @ variable flow rates |
Analytical Methodology: | Infrared Absorption Spectrophotometry |
Documentation: | 7602 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Silica, crystalline | SiO2 |
14808-60-7 |
5 |
µg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.