Method 7082

Lead by AA


This method is used for the determination of lead. The sample is collected on cellulose ester membrane filters, ashed in concentrated nitric acid and 30% hydrogen peroxide at 140 °C, and then analyzed by flame atomic absorption. The working range is 0.05 to >1 mg/m³ for a 200 L air sample. This method is applicable to elemental lead, including lead fume and all other aerosols containing lead. This is an elemental analysis, and is not compound specific. Aliquots of the samples can be analyzed separately for many additional elements. NOTE: This method combines and replaces P&CAM 173, S341 for lead. Method 7300 ( ICP-AES) and M7105 (AAS/GI) are alternate analytical techniques. M7505 is specific for lead sulfide.


(NIOSH Issue 2: 15 August 1994)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   Stable
Preservatives:   Stable
Required Preps:   Cellulose ester membrane filter (0.8µm pore size).
Collection Method:   Filter: 200L - 1500L @ 1 to 4 L/min.
Analytical Methodology:   Flame Atomic Absorption spectrophotometer
Documentation:   7082

Analyte List*

Analyte Formula CAS Number Detection Limit
Lead monoxide
OPb
1317-36-8
2.6
 µg
Lead
Pb
7439-92-1
2.6
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.