Method 6008

Stibine


This method is for the collection and analysis of stibine from wokplace air.  The sample is collected on a mercuric clhoride coated silica gel tube.  After collection, the sample is extraction with concentrated HCl and analyzed by visible spectrophotometry.  Stibine is analyzed as antimony (Sb).

The working range of the method is 0.1 to 3.5 mg/m3 (0.02 to 0.7 ppm) for a 20L sample.

NOTE:  This method revised Method S243.


(NIOSH)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   At least 7 days from sample collection to analysis if samples kept at 25°C.
Preservatives:   Keep samples at 25°C.
Required Preps:   Mercuric chloride treated silica gel tube, 1/0.5g
Collection Method:   The method recommends a 0.01 - 0.2 L/min. flow rate with a 4 - 50L sample collection volume.
Analytical Methodology:   Spectrophotometry
Documentation:   6008

Analyte List*

Analyte Formula CAS Number Detection Limit
Stibine
H3Sb
7803-52-3
0.4
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.