This method is for the determination of ribavirin. The sample is collected on glass fiber filters, desorbed in 3mL sulfuric acid (pH 2.5) with 30 minute agitation, followed by HPLC analysis with ultraviolet detection. The working range is 0.04 to 40 mg/m³ for a 50-L air sample.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | None noted in method. |
Preservatives: | Store at room temperature in the dark. |
Required Preps: | Glass fiber filter: 1-µm pore size, 37-mm diameter. |
Collection Method: | Glass fiber filter: 5L - 1000L @ 1-4L/min. |
Analytical Methodology: | HPLC/UV |
Documentation: | 5027 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Ribavirin | 36791-04-5 |
0.7 |
µg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.