This method is for the determination of hexachlorobutadiene. The sample is collected on XAD-2 sorbent tubes, desorbed in 2mL hexane with an ultrasonic bath, and analyzed by GC/ECD. The working range is 0.01 to 2.0 mg/m³ for a 100-L sample. An alternate column, DB-210, may be used with appropriate changes in instrumental conditions. Note: This revises method P&CAM 307.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | Up to 7 days @ 25 °C; 28 days @ 0 °C in dark. |
Preservatives: | Keep refrigerated, and store in the dark. |
Required Preps: | Solid sorbent tubes; XAD-2 100mg/50mg; 7 cm x 8 mm OD x 6 mm ID. |
Collection Method: | Solid sorbent tube, 1-100 L @ 0.05-0.20 L/min. |
Analytical Methodology: | GC/ECD. |
Documentation: | 2543 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Hexachlorobutadiene | C4Cl6 |
87-68-3 |
0.02 |
µg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.