This method is suitable for the determination of formaldehyde by GC/FID. The sample is collected on XAD-2, desorbed with toluene, and analyzed using GC/FID or GC/NPD. The working range is 0.24 to 16 ppm (0.3 to 20 mg/m³) for a 10-L air sample. OSHA Method 52 is similar but uses slightly larger sampling tubes.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | 3 weeks @ 25 °C |
Preservatives: | 25 °C |
Required Preps: | 10% 2-(hydroxymethyl)piperdine on XAD-2, 120 mg / 60 mg, 10 cm x 6 mm OD x 4 mm ID |
Collection Method: | Solid sorbent tube, 1-36 L @ 0.01-0.10 L/min |
Analytical Methodology: | GC/FID |
Documentation: | 2541 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Formaldehyde | CH2O |
50-00-0 |
1 |
µg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.