This method is designed for the determination of 1,3-cyclopentadiene. The sample is collected with maleic anhydride coated Chromosorb 104 sorbent tubes, desorbed with ethyl acetate, and analyzed by GC/FID. The working range is 25 to 150 ppm (67 to 400 mg/m³) for a 3 L air sample, based on sampler capacity at high relative humidity. NOTE: This method revises P&CAM 294.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | Use the coated sorbent within 20 days of preparation. Tubes are good for at least 1 week from samplling to analysis if kept at 25°C |
Preservatives: | Store at 25°C. |
Required Preps: | Solid sorbent tube: maleic anhydride coated Chromosorb 104, 100 mg/50 mg, 8.5 cm x 6mm OD x 4mm ID |
Collection Method: | Solid sorbent tube: 1 - 5 L @ 0.01 - 0.05 L/min. |
Analytical Methodology: | GC/FID |
Documentation: | 2523 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
1,3-Cyclopentadiene | C5H6 |
542-92-7 |
0.01 |
mg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.