This method is for the determination of chloroacetaldehyde. The sample is collected on silica gel sorbent tubes, desorbed in 25 mL 50% aqueous methanol for 30 minutes and analyzed by gas chromatography, with a 63Ni electron capture detector system. The working range is 0.1 to 2 ppm (0.33 to 6.3 mg/m³) for a 3 L air sample. The method is not able to distinguish between chloroacetaldehyde and the monomer or dimer hydrate forms of chloroacetaldehyde. Chloroacetaldehyde in air has also been determined by differential pulse polarography. Note: This is Method S11 in a revised format.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | At least 7 days from sampling to analysis if kept at 25°C |
Preservatives: | Store at room temperature (25°C). |
Required Preps: | Solid sorbent tube: silica gel, 520 mg/260 mg, 7cm x 6mm OD x 4mm ID |
Collection Method: | Sorbent tube: 0.3 - 16 L @ 0.05 - 0.2 L/min. |
Analytical Methodology: | GC/ECD |
Documentation: | 2015 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Chloroacetaldehyde | C2H3ClO |
107-20-0 |
0.1 |
µg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.