This method is for the determination of propylene dichloride. The sample is collected on charcoal sorbent tubes, desorbed in 1 mL 15% (v/v) acetone/cyclohexane followed by a 30 minute ultrasonic bath and then analyized by GC/FID. The working range is 0.05 to 130 ppm (0.25 to 600 mg/m³) for a 2 L air sample at about 90% relative humidity. Note: This method revises P&CAM 321.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | At least 26 days from sampling to analysis if kept at 25°C in the dark |
Preservatives: | Store at room temperature and prevent exposure to light. |
Required Preps: | Solid sorbent tube: petroleum charcoal, 100 mg/50 mg, 7cm x 6mm OD x 4mm ID |
Collection Method: | Solid sorbent tube: 0.1 - 3.5 L @ 0.01 - 0.2 L/min. |
Analytical Methodology: | GC/FID |
Documentation: | 1013 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
1,2-Dichloropropane | C3H6Cl2 |
78-87-5 |
0.1 |
µg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.