Method 1009

Vinyl Bromide by GC/FID


This method is for the analysis of vinyl bromide by GC/FID in ambient air. The working range is 0.3 to 33 ppm (1.3-145 mg/m³) for a 6 L air sample. The sample is collected on charcoal sorbent tubes and analyzed by GC/FID after desoprtion in ethanol, followed by ultrasonic agitation. Note: This is Method P&CAM 349 in a revised format.


(NIOSH Issue 2: 15 August 1994)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   At least 14 days from collection to analysis if kept at 25°C
Preservatives:   Store at 25°C.
Required Preps:   Solid sorbent tube: 2 sections of coconut shell charcoal, 400 mg / 200 mg, 10cm x 8mm OD x 6mm ID
Collection Method:   Solid sorbent tube: 2 - 10 L @ 0.01 - 0.2 L/min.
Analytical Methodology:   GC/FID
Documentation:   1009

Analyte List*

Analyte Formula CAS Number Detection Limit
Vinyl Bromide
C2H3Br
593-60-2
3
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.