This method is used for the determination of chloroprene.The sample is collected on a charcoal sorbent tube, desorbed in carbon disulfide, and analyzed using GC/FID. The method is sensitive enough to determine concentrations as low as 12 ppm in 15 minute samples taken at 0.2 L/min. Note: This method is S112 with an improved GC column in a new format.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | Up to 8 days from sampling to analysis if kept at 25°C |
Preservatives: | Store at 25 °C. |
Required Preps: | Solid sorbent tube: coconut shell charcoal, 100 mg/50 mg, 7cm x 6mm OD x 4mm ID |
Collection Method: | Solid sorbent tube: 1.5 - 8 L @ 0.01 - 0.1 L/min. |
Analytical Methodology: | GC/FID |
Documentation: | 1002 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Chloroprene | C4H5Cl |
126-99-8 |
0.03 |
mg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.