This method is for the determination of allyl chloride. The sample is collected on charcoal sorbent tube, desorbed with benzene, and analyzed using GC/FID. This method is applicable to short term samples taken at 1L/min. The working range is 0.16 to 3 ppm (0.5 to 10mg/m³) for a 100 L air sample. Note: This is method S116 in revised format.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | At least 6 weeks from sampling to analysis if kept at 25 °C. |
Preservatives: | 25 °C. |
Required Preps: | Solid sorbent tube: coconut shell charcoal, 100 mg/50 mg, 7cm x 6mm OD x 4mm ID |
Collection Method: | Solid sorbent tube: 16 -100 L @ 0.01 - 1 L/min. |
Analytical Methodology: | GC/FID |
Documentation: | 1000 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Allyl chloride | C3H5Cl |
107-05-1 |
0.01 |
mg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.