This method measures the mass concentration of any non-volatile respirable dust. Besides inert dusts, the method has been recommended for respirable coal dust. The method is biased in light of the recently adopted international definition of respirable dust (e.g., ˜+7% bias for non-diesel, coal mine dust). The working range is 0.5 to 10 mg/m³ for a 200-L air sample. NOTE: This method is based on and replaces Sampling Data Sheet #29.02.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | Indefinite |
Preservatives: | None |
Required Preps: | Tared Filter 5 µm PVC membrane. |
Collection Method: | Cyclone (10 mm or HD Cyclone) and filter; 20-400 L@ 5 mg/m³. Flow Rate: 1.7 L/min Nylon Cyclone, 2.2 L/min HD Cyclone |
Analytical Methodology: | Gravimetric (Filter Weight) |
Documentation: | 0600 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Respirable Particulates | 0.03 |
mg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.