Method PV2141

Triethanolamine


This method is for the collection and analysis of triethanolamine from workplace air. The sample is collected on closed-face 37 mm glass fiber filter cassette. After collection, the sample is desorbed with acetone and analyzed by GC/FID (flame ionization detector).

NOTE: This is a partially validated method.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   The method states recoveries of the analyte of 95% or higher when stored for 14 days at ambient or refrigerated temperatures.
Preservatives:   Store samples at ambient or refrigerated temperatures.
Required Preps:   37 mm glass fiber filter, SKC 225-709
Collection Method:   The method recommends a 1 L/min. flow rate with a 100L sample volume.
Analytical Methodology:   GC/FID
Documentation:   PV2141

Analyte List*

Analyte Formula CAS Number Detection Limit
Triethanolamine
C6H15NO3
102-71-6
10
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.