Method PV2130

Benzophenone


This method is for the collection and analysis of benzophenone from workplace air.  The sample is collected on a Chromosorb 106 tube.  After collection, the sample is extracted with a solution of carbon disulfide/N, N-dimethylformamide (99:1) and analyzed by GC/FID (flame ionization detector).

NOTE:  Ths is a partically validated method.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   The analyte recoveries were above 97% when stored at ambient and refrigerated temperatures for 14 days.
Preservatives:   Store at ambient or refrigerated temperatures.
Required Preps:   Chromosorb 106, 50/100mg (SKC 226-110)
Collection Method:   The method recommends a 0.2 L/min. flow rate and a 48L sample collection volume.
Analytical Methodology:   GC/FID
Documentation:   PV2130

Analyte List*

Analyte Formula CAS Number Detection Limit
Benzophenone
C13H10O
119-61-9
0.89
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.