This method is for the determination of total residue from drinking, surface, and saline waters, domestic and industrial wastes. A well mixed aliquot of the sample is quantitatively transferred to a pre-weighed evaporating dish and evaporated to dryness at 103-105°C. Total residue is defined as the sum of the homogenous suspended and dissolved materials in a sample.
The practical range of the method is from 10 mg/L to 20,000 mg/L.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | 7 days from sample collection to analysis. |
Preservatives: | Store sample at 4°C. |
Required Preps: | 250mL Glass or plastic bottle with Teflon-lined lid |
Collection Method: | Grab sampling |
Analytical Methodology: | Gravimetric |
Documentation: | 160.3 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Total Solids | 10 |
mg/L |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.