Method 94

Methyl methacrylate by GC/FID


This method describes the sample collection and analysis of methyl methacrylate. Samples are collected by drawing air through glass sampling tubes containing coconut shell charcoal coated with 4-tert-butylcatechol. Samples are desorbed with toluene and analyzed by GC using a flame ionization detector.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   Up to 15 days.
Preservatives:   Samples should be stored at reduced temperature when not in transit.
Required Preps:   Solid sorbent tubes: coconut shell carbon coated with TBC, 10% by weight. 110mg/55mg, 7cm x 6mmOD x 4mmID.
Collection Method:   Solid sorbent tube: 3 L at 0.05 L/min.
Analytical Methodology:   GC/FID.
Documentation:   094

Analyte List*

Analyte Formula CAS Number Detection Limit
Methyl methacrylate
C5H8O2
80-62-6
1.852
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.