Method 86

Maleic anhydride by HPLC/UV


This method describes the sample collection and analysis of maleic anhydride. Samples are collected by drawing air through glass fiber filters coated with 2 mg of 3,4-dimethoxybenzylamine (veratrylamine). Samples are extracted with 90:10 (v/v) acetonitrile/dimethylsulfoxide and analyzed by HPLC using a UV detector.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   Submit the samples for analysis as soon as possible after sampling.
Preservatives:   Store samples in a refrigerator upon receipt at the laboratory. If delay is unavoidable, store the samples in a refrigerator.
Required Preps:   Two glass fiber filters, each coated with 2 mg of veratrylamine contained in a three-piece polystyrene cassette.
Collection Method:   Treated glass fiber filters: 60 L at 0.5 L/min.
Analytical Methodology:   HPLC/UV
Documentation:   086

Analyte List*

Analyte Formula CAS Number Detection Limit
Maleic Anhydride
C4H2O3
108-31-6
1.99
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.