Method 74

Aldicarb


This method describes the sample collection and analysis of aldicarb (temik). Samples are collected by drawing known volumes of air through OSHA Versatile Samplers containing a glass fiber filter and two sections of XAD-2 adsorbent (OVS-2). Samples are extracted/desorbed with ace tone and analyzed by GC using a nitrogen/phosphorus detector.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   At least 17 days.
Preservatives:   Store at room temperature.
Required Preps:   Samples are collected with glass OVS-2 tubes,13-mm OD tubes that are tapered to 6-mm OD. The tubes are packed with a 140-mg back section and a 270-mg front section of cleaned XAD-2 resin along with a 13-mm diameter glass fiber filter.
Collection Method:   Solid sorbent XAD-2 tubes with glass fiber filters: 480 L at 1.0 L/min
Analytical Methodology:   GC/NPD.
Documentation:   074

Analyte List*

Analyte Formula CAS Number Detection Limit
Aldicarb (Temic)
C7H14N2O2S
116-06-3
120.5
 ng

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.