Method 70

Pyrethrum by GC/ECD


This method describes the sample collection and analysis of pyrethrum. Samples are collected by drawing known volumes of air through OSHA Versatile Sampler tubes containing a glass fiber filter and two sections of XAD-2 adsorbent (OVS-2) tubes. The OVS-2 is a specially prepared glass tube containing a glass fiber filter and XAD-2 resin which will collect both vapors and aerosols. Samples are extracted/desorbed with toluene and analyzed by GC using an electron capture detector.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   Recovery remained above 95.6% after 15 days of storage at ambient temperature (22°C) or refrigeration at -14°C
Preservatives:   Keep refrigerated or at ambient temperature.
Required Preps:   Samples are collected with glass OVS-2 tubes,13-mm OD tubes that are tapered to 6-mm OD. The tubes are packed with a 140-mg back section and a 270-mg front section of cleaned XAD-2 resin along with a 13-mm diameter glass fiber filter.
Collection Method:   Solid sorbent XAD-2 tubes with glass fiber filters: 60L @ 1L/min.
Analytical Methodology:   GC-ECD
Documentation:   070

Analyte List*

Analyte Formula CAS Number Detection Limit
Pyrethrum
8003-34-7
2.99
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.