This method describes the sample collection and analysis of acetaldehyde. Air samples are collected by drawing known volumes of air through 8mm OD sampling tubes containing XAD-2 adsorbent which has been coated with 2-(hydroxymethyl) piperidine. Each sampling tube contains a 450-mg sampling section and a 225-mg backup section. The samples are desorbed with toluene and the derivatized acetaldehyde is analyzed by gas chromatography using a nitrogen selective detector.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | At least 23 days. |
Preservatives: | Store at room temperature or keep refrigerated. |
Required Preps: | Silane-treated glass tubes packed with pretreated XAD-2 adsorbent coated with 2-HMP; 450mg/225mg 11cm x 8mmOD x 6mmID. |
Collection Method: | Treated XAD-2 tubes: 3 L at 0.05 L/min. |
Analytical Methodology: | GC/NPD |
Documentation: | 068 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Acetaldehyde | C2H4O |
75-07-0 |
3.14 |
µg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.