This method is for the sample collection and analysis of technical grade chlordane.  The sample is collected and an OVS-2 tube which contains a glass fiber filter and two sections of XAD-2.  After collection the sample is extracted with toluene and analyzed by GC/ECD (electron capture detector). 


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   Recovery of chlordane remained above 95.6% after 15 days of storage at ambient temperature (22°C).
Preservatives:   Store samples at 22°C.
Required Preps:   OVS-2 tube, 140/270 mg (SKC 226-30-16)
Collection Method:   The method recommends a 1 L/min. flow rate and a 480L sample collection volume.
Analytical Methodology:   GC/ECD
Documentation:   067

Analyte List*

Analyte Formula CAS Number Detection Limit
Chlordane
C10H6Cl8
57-74-9
30.5
 ng

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.