This method describes the sample collection and analysis of acrylonitrile. Samples are collected on charcoal, desorbed with acetone, and analyzed by gas chromatography using a nitrogen/phosphorus detector.
No Obligation Quotation for Analytical Services
If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Hold Times, Preservatives, Preps, Collection, Analytical & Documentation | |
---|---|
Holding Time: | At least 15 days. |
Preservatives: | Store at room temperature. |
Required Preps: | Solid sorbent tubes: Coconut shell charcoal tubes, 100mg/50mg (glass tubes 7-cm x 6-mm OD x 4-mm) . |
Collection Method: | Solid sorbent tubes: 20 L at 0.2 L/min. |
Analytical Methodology: | GC/NPD. |
Documentation: | 037 |
Analyte | Formula | CAS Number | Detection Limit | |
---|---|---|---|---|
Acrylonitrile | C3H3N |
107-13-1 |
0.51 |
µg |
* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.