Method PV2096

N,N-Dimethylethylamine


This method is for the determination of N,N-dimethylethylmine.  The sample is collected using a jumbo alumina tube.  After sample collection the tube is desorbed using deionized water and analyzed using GC/NPD (nitrogen-phosphorus detector). 

The capacity of the different lots of alumina tubes to collect the analyte varied greatly.  The same tube lot should be used for the entire project if possible.

Note that this method is only partially validated.

 


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   A stability study showed recoveries above 95% after 14 days when kept at room temperature.
Preservatives:   None specified in the method.
Required Preps:   Jumbo alumina tube (SKC 226-18)
Collection Method:   The method recommends a flow rate of 0.1 L/min. with collection volume of 4L.
Analytical Methodology:   GC/NPD
Documentation:   PV2096

Analyte List*

Analyte Formula CAS Number Detection Limit
N,N-Dimethylethylamine
C4H11N
598-56-1
0.3
 ppm

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.