Method 95

Ethylene thiourea by HPLC-UV


This method describes the sample collection and analysis of ethylene thiourea. Samples are collected open-face by drawing air through a four-piece sampling cassette containing two glass fiber filters. Samples are extracted with water and analyzed by HPLC using a UV detector.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   Up to 15 days.
Preservatives:   Samples should be stored in a refrigerator when not in transit.
Required Preps:   A four-piece polystyrene cassette containing two glass fiber filters.
Collection Method:   Glass fiber filters: 480 L at 2.0 L/min.
Analytical Methodology:   HPLC/UV.
Documentation:   095

Analyte List*

Analyte Formula CAS Number Detection Limit
2-Imidazolidinethione
C3H6N2S
96-45-7
0.666
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.